7. MEASUREMENT
The instrument ALMEMO
®
5590-2 provides the following options for the
acquisition of measuring data:
1. Continuous measurement of a selectable meas. point, see manual 6.4
Output of measuring data to the analogue output, see manual 5.1.1.
2. Single measuring point scan, see manual 6.5.1.1
3. Cyclic measuring point scan, see manual 6.5.1.2
4. Continuous measuring point scan, see manual 6.5.1.3
Total Clearing of all Measured Values
Previous measuring data should be cleared before a measurement. Max, min,
and average values of all channels and the memory can be cleared with the
rotary switch in position MEMORY, by using the keys
ENTER
,
ENTER
,
CLEAR
.
Key:
flashes
(
clear memory)
ENTER
Clr SP
flashes
(
clear memory and meas. values)
ENTER
Clr All
cleared (not with other key)
CLEAR
-
-
-
For automatic clearing on each START, see manual 6.10.13.2.
7.1 Continuous Measurement of a Measuring Point
As long as no cycle and no continuous measuring point scan have been
programmed (e.g. after a reinitialisation, see 3.4) only the measured value of a
selected measuring point, which is at first M0, is continuously acquired with the
specified conversion rate (see 7.3.3) (optimal for analogue output).
7.1.1 Measured Value, Selecting a Measuring Point
With switch position
MEAS.VALUE
the display provides the following information.
In the 1st line a combination of device address 00 (s.8.2) and meas. point 12 is
displayed. The short name of the meas. range follows (s.6.2) if a measuring
point designation has not been entered (s.6.4). At the end the measured value
is displayed including the dimension. In the 2nd line the version of the selected
module or a programmed module designation (s.5.5) will be displayed.
Function MEAS.VALUE:
MEAS.VALUE
0012: NiCr 123.45°C
A5590-2 5.50