6.3.1.2 Changing Gas Sensor Type
Figure 58: Different Gas Sensor Warning
A warning screen is displayed when the detector type is changed. This is to prevent accidental detector
type changes. A gas sensor type change results in the loss of all gas specific information from the
display, including:
Setpoints
Healthy region setting
Hysteresis setting
Last response test target
In addition, the data listed under Section 0 is removed.
After acknowledging this screen by pressing the centre magnetic button, the display will either:
Begin the detector image update process, refer to Section 6.3.1.4, or
Show the detector warmup screen, refer to Section 6.2.2.