4.4.2.2 GG2 Remote Cable Assembly
The GG2 Remote Cable Assembly is mounted hanging down from a flat surface using the three screw
tabs. The custom plug interface allows subsequent mounting of compatible modules. Refer to Figure 9
for mounting hole locations. The three mounting holes are spaced 50mm on centre.
Connecting and Disconnecting a GG2 Detector
The process for connecting and disconnecting the GG2 Detector is outline in Figure 10 and detailed in
the following subsections. The detector must be isolated or in sleep mode prior to
connection/disconnection. Refer to Section 6.3 for more information.
Figure 10: Connecting/Disconnection Detector
4.4.3.1 Disconnecting Detector
1. Release the retainer clip by pulling firmly away from the display. Ensure the detector is supported
at all times to prevent potential damage.
2. After releasing the retainer clip, release the detector by pulling the detector firmly downwards.
4
1
2
3