CAUTION
The voltage rating of the DC Analyzer is 10–28V.
•DC input has to be an approved Class 2 or limited energy circuit for DC power
•Voltages outside this range may cause the Analyzer to malfunction.
The voltage rating of the AC Analyzer is 100 to 240VAC at 50/60Hz ± 10%.
•Any AC voltages outside this range may cause the Analyzer to malfunction
Any use of this equipment in a manner not specified in this manual or approved AMI documentation
may impair the protection provided by the equipment.
Toute utilisation de cet équipement d'une manière non spécifiée dans ce manuel ou dans la documen-
tation AMI approuvée peut altérer la protection fournie par l'équipement.
A SEAL SHALL BE INSTALLED WITHIN 50 mm OF THE ENCLOSURE.
UN SCELLEMENT DOIT ETRE INSTALLE A MOINS DE 50 mm DU BOITIER.
WARNING
SUBSTITUTION OF COMPONENTS MAY IMPAIR INTRINSIC SAFETY.
LE REMPLACEMENT DE COMPOSANTS PEUT COMPROMETTRE LA SECURITE INTRINSEQUE.
WARNING
Enclosure materials contain a light metal content of over 10% Aluminum and pose a potential impact
spark ignition hazard. The end user shall carry out a risk assessment prior to installation in an EPL Ga
environment and shall only install the equipment where the risk of impact has been considered to be
negligible.
Les matériaux de boîtier contiennent une teneur en métaux légers de plus de 10% d'aluminium et con-
stituent un risque potentiel d'inflammation. L'utilisateur final doit procéder à une évaluation des risques
avant de l'installer dans un environnement EPL Ga et ne doit installer le matériel que dans les cas où le
risque d'impact a été considéré comme négligeable.
WARNING