![AMI The Barracuda 4010LX Скачать руководство пользователя страница 59](http://html.mh-extra.com/html/ami/the-barracuda-4010lx/the-barracuda-4010lx_operators-manual_2932178059.webp)
Register Number of
Register
Used
Variable
Name
Description
Type
Comment
61
8
A0RJ0
Analyzer Type
String
69
1
A0RJ1
Analyzer Configuration
16-bit Unsigned Integer
70
8
A0RL0
Analyzer Serial Number
String
78
8
A0RL1
Analyzer Tracking Number String
86
8
A0RL2
Analyzer User ID
String
94
8
A0RL3
Analyzer Laser S/N
String
102
10
A0RM0
Latest Startup Info
String
112
2
A0RN0
Write Com ID
String
114
1
A0RN1
Modbus ID
16-bit Unsigned Integer
115
10
A0RO0
Low Power Event
String
125
1
A0RP0
RTC Seconds
16-bit Unsigned Integer
126
1
A0RP1
RTC Minutes
16-bit Unsigned Integer
127
1
A0RP2
RTC Hours
16-bit Unsigned Integer
128
1
A0RP3
RTC DOW
16-bit Unsigned Integer
129
1
A0RP4
RTC DOM
16-bit Unsigned Integer
130
1
A0RP5
RTC Month
16-bit Unsigned Integer
131
1
A0RP6
RTC Year
16-bit Unsigned Integer
132
1
A0RP7
Log Interval
16-bit Unsigned Integer
133
1
A0RT0
Block Tempe26
16-bit Unsigned Integer
134
1
A0RT2
Pressure x 10000 (bar)
16-bit Unsigned Integer
135
1
A0RT3
Power Voltage x 100
16-bit Unsigned Integer
136
1
A0RU0
Hours of Operations
16-bit Unsigned Integer
137
1
A0RW0
Alarm Pulse Time
16-bit Unsigned Integer
138
1
A0RX0
Delay on for Alarm 1
16-bit Unsigned Integer
139
1
A0RY0
Delay on for Alarm 2
16-bit Unsigned Integer
140
1
A0RZ0
Alarm Hold of Time
16-bit Unsigned Integer
Table I: Holding Registers for BARRACUDA MODEL 4010LX (continued)