5.
Verify the operating voltage of your Analyzer and the correct power requirements before
you continue.
6.
Make sure the power source has been turned-off before you begin installing wiring.
3.
Remove the explosion-proof cover by rotating it counterclockwise.
Note: A white sheet metal panel inside the explosion-proof housing indicates the
Analyzer has been built for use with DC power, while a black sheet metal panel indicates
AC power.
4.
Then remove the Terminal Cover to access the electrical connections.
Terminal Cover
DC Version with Terminal Cover
and white information panel
Terminal Cover
AC Version with Terminal Cover
and black information panel
IMPORTANT
: When attaching wiring to the green terminal connectors, use either solid wire or
stranded wire with wire ferrule(s) attached. Verify no loose strands are visible after
installation of wire ferrule(s).
•
The green terminal block connectors are combination connectors, which allows you to unplug
the connector during the wiring process. Combination connectors can accommodate
12–24 AWG wire sizes for your electrical connections
12