![AMI The Barracuda 4010LX Скачать руководство пользователя страница 38](http://html.mh-extra.com/html/ami/the-barracuda-4010lx/the-barracuda-4010lx_operators-manual_2932178038.webp)
Resolution
:
First, check to verify that the laser power is present. The blue waveform represents the laser power of
THE
BARRACUDA MODEL 4010LX
. If its pattern appears, as shown above, it means that the laser is func-
tioning properly.
Then check your Sample Inlet from the pipeline to
THE BARRACUDA
, making sure that all connections are
secured and have no leaks. Also, purge the Sample Line for a few minutes before resuming taking moisture
measurements.
Graph with Shifted Waveforms Out of Tuning Range
Potential Issue
:
Over time, laser-based Moisture Analyzers eventually see a gradual movement of their signature peaks on
the x-axis of the measurement waveform. This shift can potentially affect the moisture measurement.
Resolution
:
Use
SMART REALIGNMENT
™
on the
COMMAND CENTER
to realign the critical H
2
O and CH
4
peaks.
CH
4
Peak
has slightly moved
from the frequency range that
THE BARRACUDA
targets
Moisture Peak
has also
moved from the frequency
range that the Analyzer targets
37