Analog Programming (APG) and Isolated Analog Programming (ISOL)
4-32
M370185-01
Query for ISOL Current Input Level
To query for ISOL current input level from non-isolated input:
1. Turn the 9-position mode control to the CAP position or press the
rotary Adjust/Enter control if the control knob is already at the CAP
position.
CAPr
is displayed on the output voltage display.
2. If the mode is currently
nonE
then select any other ISOL mode. If
this step is done, be sure to restore the ISOL setting to
nonE
when
the operation is complete. Otherwise, do not rotate the rotary
Adjust/Enter control to keep the current ISOL mode.
3. Press the rotary Adjust/Enter control to select the ISOL mode.
The output voltage display should read
LE C
and the output current
display will display ISOL level.
4. Press the rotary Adjust/Enter control to return to normal mode.
5. If the ISOL mode was modified in step 2, execute the procedure to
deactivate the ISOL voltage mode.
Quick Tip
Remote operation
uses these SCPI
commands. See
page 5–1.
The SCPI command for these instructions are:
[:]SYSTem[<channel>]:REMote:SOURce:CURRent <?|LOCal
|AVOLtage|IAVoltage|ARESistive|IAResistive>
[:]SYSTem[<channel>]:COMMunicate:APRogram:LEVel:CURRent
[:ISOlated]
[:]MEASure[<channel>][:SCALar]:APRogram:CURRent
:ISOlated[:DC]?
Содержание Soresen XTR 100-8.5
Страница 2: ......
Страница 16: ...xiv ...
Страница 28: ...1 8 ...
Страница 50: ...Local Operation 3 8 M370185 01 Figure 3 3 Front Panel Menu System ...
Страница 93: ...Using Multiple Power Supplies M370185 01 3 51 3 ...
Страница 94: ...Local Operation 3 52 M370185 01 ...
Страница 128: ...Analog Programming APG and Isolated Analog Programming ISOL 4 34 M370185 01 ...
Страница 198: ...Remote Operation 5 70 M370185 01 ...
Страница 261: ...SCPI Command Tree M370185 01 A 35 A ...
Страница 262: ...A 36 ...
Страница 279: ...Index M370185 01 IX 3 described 3 7 3 36 number available 3 36 stored values 3 36 ...
Страница 280: ...IX 4 ...