Sorensen SGI Series
Operation
M550221-01 Rev U
3-43
3.10
Remote Current Programming
Remote current programming is summed with the front panel or digital setting;
see Section 3.9. Remote current programming is used for applications that
require the output current be programmed (controlled) from a remote
instrument. An external resistance or external voltage source may be used as
a programming device. When using remote current programming, a shielded,
twisted-pair cable is recommended to prevent noise interference to
programming signals.
3.10.1
Remote Current Programming by Resistance
The resistance-programming coefficient for output current is
(100% rated output current) /
5 kΩ, with input at Pin J1-22 (IP RES) and return
to Pin J1-23 (IP RTN). An internal current source,
factory-set at 1 mA, from
Pin J1-22 (IP RES) is utilized to drive the resistance. This produces a transfer
function for output current, as follows:
Iout = R * (100% rated output current) /
5 kΩ), with R in ohms.
If multiple switches or relays are used to select resistors to program different
current levels, make-before-break contacts are recommended.
Note:
If an external resistance is used for remote programming, the current
programming return Pin J1-23 (IP RTN), must be connected directly to, or
within ±3 V, of the circuit common, Pins J1-6 and J1-24. See Figure 3-14 for
connection requirements.
Figure 3-14. Remote Current Programming Using Resistance
Содержание Sorensen SGI Series
Страница 1: ...M550221 01 Rev U www programmablepower com SGI Series DC Power Supplies Operation Manual...
Страница 2: ......
Страница 34: ...Installation Sorensen SGI Series 2 12 M550221 01 Rev U Figure 2 2 Installation Drawing 3U Models 10V 30V...
Страница 35: ...Sorensen SGI Series Installation M550221 01 Rev U 2 13 Figure 2 3 Installation Drawing 3U Models 40V 600V...
Страница 36: ...Installation Sorensen SGI Series 2 14 M550221 01 Rev U Figure 2 4 Installation Drawing 3U Models 800V and 1000V...
Страница 37: ...Sorensen SGI Series Installation M550221 01 Rev U 2 15 Figure 2 5 Installation Drawing 6U Models 20kW 30kW...
Страница 38: ...Installation Sorensen SGI Series 2 16 M550221 01 Rev U Figure 2 6 Rear Panel Standard 3U Models 10V 30V...
Страница 39: ...Sorensen SGI Series Installation M550221 01 Rev U 2 17 Figure 2 7 Rear Panel GPIB Option 3U Models 10V 30V...
Страница 40: ...Installation Sorensen SGI Series 2 18 M550221 01 Rev U Figure 2 8 Rear Panel Ethernet Option 3U Models 10V 30V...
Страница 41: ...Sorensen SGI Series Installation M550221 01 Rev U 2 19 Figure 2 9 Rear Panel Standard 3U Models 40V 600V...
Страница 42: ...Installation Sorensen SGI Series 2 20 M550221 01 Rev U Figure 2 10 Rear Panel GPIB Option 3U Models 40V 600V...
Страница 44: ...Installation Sorensen SGI Series 2 22 M550221 01 Rev U Figure 2 12 Rear Panel Standard 6U Models 20kW 30 kW...
Страница 45: ...Sorensen SGI Series Installation M550221 01 Rev U 2 23 Figure 2 13 Rear Panel GPIB Option 6U Models 20kW 30 kW...
Страница 48: ...Installation Sorensen SGI Series 2 26 M550221 01 Rev U This page intentionally left blank...
Страница 103: ...Sorensen SGI Series Operation M550221 01 Rev U 3 55 Figure 3 26 Parallel Connection and Remote Sense...
Страница 116: ...Maintenance Sorensen SGI Series 5 4 M550221 01 Rev U This page intentionally left blank...
Страница 120: ...Index Sorensen SGI Series Index 4 M550221 01 Rev U This page intentionally left blank...