3.15
Floating and Polarized Output
The SGI Series supply can be set up for a Positive or Negative supply, as well
as standard operation as a floating output supply.
F
LOATING
O
UTPUT
The output terminals are normally floating from chassis ground. No extra
steps or connections are required for a floating output.
P
OSITIVE
S
UPPLY
S
ETUP
Attach the negative output terminal to the supply chassis. The output
reference is now chassis ground. When the output voltage is set or
programmed, the supply will output a positive potential from chassis ground.
N
EGATIVE
S
UPPLY
S
ETUP
Attach the Positive output terminal to the supply chassis. The output
reference is now chassis ground. When the output voltage is set or
programmed, the supply will output a negative potential from chassis
ground.
CAUTION!
The negative output terminal may be floated up to ±300V (PK), maximum,
with respect to chassis ground. Exceeding the limit will be detected as a
fault by a protective supervisory monitor and shutdown of the output will
be executed; this condition will be latched, requiring reset to resume
normal operation.
CAUTION!
Floating the negative output terminal subjects the internal control circuitry of
the power supply to the same potential as present at the negative output
terminal. In a unit with the standard Non-Isolated Analog Interface, the signals
of control connector, J1, would float at the same potential as the negative
output terminal. Damage might occur if the signals of the Non-Isolated Analog
control connector are connected to an external ground referenced device, due
to unintentional ground loop currents that this connection could generate. To
correct ground loop problems, it is advised to use the optional Isolated Analog
Interface in order to isolate the external signals from the internal control
circuitry of the supply. Refer to the Section 1.2.2 for additional information.
Содержание Sorensen SGI Series
Страница 1: ...M550221 01 Rev U www programmablepower com SGI Series DC Power Supplies Operation Manual...
Страница 2: ......
Страница 34: ...Installation Sorensen SGI Series 2 12 M550221 01 Rev U Figure 2 2 Installation Drawing 3U Models 10V 30V...
Страница 35: ...Sorensen SGI Series Installation M550221 01 Rev U 2 13 Figure 2 3 Installation Drawing 3U Models 40V 600V...
Страница 36: ...Installation Sorensen SGI Series 2 14 M550221 01 Rev U Figure 2 4 Installation Drawing 3U Models 800V and 1000V...
Страница 37: ...Sorensen SGI Series Installation M550221 01 Rev U 2 15 Figure 2 5 Installation Drawing 6U Models 20kW 30kW...
Страница 38: ...Installation Sorensen SGI Series 2 16 M550221 01 Rev U Figure 2 6 Rear Panel Standard 3U Models 10V 30V...
Страница 39: ...Sorensen SGI Series Installation M550221 01 Rev U 2 17 Figure 2 7 Rear Panel GPIB Option 3U Models 10V 30V...
Страница 40: ...Installation Sorensen SGI Series 2 18 M550221 01 Rev U Figure 2 8 Rear Panel Ethernet Option 3U Models 10V 30V...
Страница 41: ...Sorensen SGI Series Installation M550221 01 Rev U 2 19 Figure 2 9 Rear Panel Standard 3U Models 40V 600V...
Страница 42: ...Installation Sorensen SGI Series 2 20 M550221 01 Rev U Figure 2 10 Rear Panel GPIB Option 3U Models 40V 600V...
Страница 44: ...Installation Sorensen SGI Series 2 22 M550221 01 Rev U Figure 2 12 Rear Panel Standard 6U Models 20kW 30 kW...
Страница 45: ...Sorensen SGI Series Installation M550221 01 Rev U 2 23 Figure 2 13 Rear Panel GPIB Option 6U Models 20kW 30 kW...
Страница 48: ...Installation Sorensen SGI Series 2 26 M550221 01 Rev U This page intentionally left blank...
Страница 103: ...Sorensen SGI Series Operation M550221 01 Rev U 3 55 Figure 3 26 Parallel Connection and Remote Sense...
Страница 116: ...Maintenance Sorensen SGI Series 5 4 M550221 01 Rev U This page intentionally left blank...
Страница 120: ...Index Sorensen SGI Series Index 4 M550221 01 Rev U This page intentionally left blank...