digiBASE
®
PMT Base with Integrated Bias Supply, Preamplifier and Digital Multichannel Analyzer
931003E / 0620
Example Responses:
$F0OFF<CR>
Gate disabled.
$F1COIN<CR>
Gate set to coincidence mode. Low during peak detect
rejects conversions.
$F2ENABLE<CR>
Gate functions as ENABLE. Stops clocks and conversions
when low.
$F3EVENT<CR>
Gate functions as event counter.
SHOW_HV
Reports the current high voltage and the status of the high voltage power supply in the form
$Dvvvvvsssssccc.
Where vvvvv represents the current output voltage if the high voltage is enabled, or the rear-
panel high voltage setting if the high voltage is disabled. sssss represents the status of the
high voltage bias supply as a 16-bit decimal number with the following bit definitions:
Bit 0 (LSB):
Bias supply polarity (0=positive, 1=negative).
Bit 1:
Bias supply overload (0=overload, 1=normal).
Bit 2:
High voltage enabled (0=disabled, 1=enabled).
Example Responses:
$D0100000003077<CR>
1000 V, negative, not overloaded, disabled.
$D0100000002076<CR>
1000 V, positive, not overloaded, disabled.
$D0100000007082<CR>
1000 V, negative, not overloaded, enabled.
SHOW_HV_ACTUAL
Returns the value of HV actually on the detector.
SHOW_HV_POLARITY
This returns the HV polarity setting as a $F record (always positive for the digiBASE).
Response:
$IPOS<CR>
The HV is positive.
SHOW_HV_TARGET
Under normal operation, the HV will go to this value when the HV is enabled. Reports the
current HV setting (see SET_HV) and the status of the HV power supply in the form:
$Dvvvvvsssssccc<CR>
where
vvvvv
represents the HV setting, and
sssss
represents the status of the HV bias supply
as a 16-bit decimal number with the following bit definitions:
Bit 0 (LSB):
Bias supply polarity (0=positive, 1=negative).
Bit 1:
Bias supply overload (0=overload, 1=normal).
Bit 2:
HV enabled (0=disabled, 1=enabled).
36