digiBASE
®
PMT Base with Integrated Bias Supply, Preamplifier and Digital Multichannel Analyzer
931003E / 0620
Responses:
$C00050092
The LLD is 50.
SHOW_MODE
Reports mode of operation (List, PHA, or Sample [InSight]).
Response:
$FPHA<CR>
PHA mode.
$FLIS<CR>
List mode.
$FSAM<CR>
Sample (InSight) mode.
SHOW_MONI_LABEL
num
Returns with an ASCII string label designation for the monitor num.
SHOW_MONI_MAX
Returns with the number of monitors available for viewing.
SHOW_MONI_VALUE
num
Returns with an ASCII representation of the value for the monitor.
SHOW_NaI
Indicates whether MCB is in NaI Mode. Responses are true and false.
Responses:
$IT<CR>
digiBASE is always in NaI mode.
SHOW_NEXT
Used in conjunction with the SHOW_ROI command, SHOW_NEXT reports the next
continuous group of channels that have the ROI flag set. The response is of the form
$Dsssssnnnnnccc<CR>
where sssss represents an integer number that is the number of the first channel of the “next”
group of channels that all have their ROI bit set, and nnnnn represents an integer number that
is the number of channels in the group. If no more channels have their ROI bit set,
SHOW_NEXT returns a first channel of 0 and a number of channels of 0. The SHOW_ROI
command is used to report the “first” group of channels that all have their ROI bit set.
Example Responses:
$D0100000050078<CR>
Next ROI group starts at channel 1000 and is 50 channels
long.
$D0215000150086<CR>
Next ROI group starts at channel 2150 and is 150 channels
long.
$D0000000000072<CR>
No other ROI groups to report.
38
Содержание ORTEC digiBASE
Страница 6: ...vi...