digiBASE
®
PMT Base with Integrated Bias Supply, Preamplifier and Digital Multichannel Analyzer
931003E / 0620
LIST_GATE
Lists the various ways the GATE (ENABLE on the digiBASE) can be used.
Response:
GATE Off Coincidence Enable Event
LIST_HV
Lists the valid high-voltage settings. digiBASE accepts HV values between 0 and +1200 V
in 960 steps (1.25 V/step).
LIST_ROI_SAMPLE
Lists the various status bits that can be displayed in InSight mode.
Response:
ROI_SAMP RESV RESV RESV LLD BUSY ULD RESV PKDET
LIST_SHAP
Displays the valid shaping times in units of 10 ns.
Response:
SHAPING 75 100 125 150 175 200
LIST_TRIG_SAMPLE
Lists the various trigger sources in InSight mode.
Response:
TRIG_SAMP Enable PeakDet LLD Random
LIST_ZERO_ADJUST
Displays the range of zero-stabilizer adjustment that can be returned with the
SHOW_ZERO_ADJUST command. A value of
!
100 corresponds to the internal value of
!
100 and +100 corresponds to +100.
Response:
ZERO_ADJU
!
100 100
!
100 100
SAMPLE
Causes a waveform to be captured in InSight mode.
Response:
%000128080
No waveform was available
%000000069
New waveform was successfully captured
SET_DEVICE
Included for backward compatibility. This command executes SET_WINDOW, resetting the
window of interest to the maximum number of channels.
30