digiBASE
®
PMT Base with Integrated Bias Supply, Preamplifier and Digital Multichannel Analyzer
931003E / 0620
Figure 6
. Change MCB Number or Description
Make the desired changes and click
Close
. Any changes you have made to an ID number or
description will then be written back to the corresponding MCB.
If a modified description has already been applied to a particular instrument, you can restore the
default description by deleting the entry in the
Description
field and re-running MCB Config-
uration. After MCB Configuration runs, the default description will be displayed.
2.3.5. Attaching More Than One digiBASE to the Computer
Once the drivers have been installed for one digiBASE, adding subsequent units is simple.
REMINDER
Be sure to run MCB Configuration any time you add new digiBASEs (or other
ORTEC MCBs) to your system,
or when you move an instrument from one USB
port to another
.
2.3.6. Connecting to and Disconnecting from the Computer
The USB connection allows you to connect digiBASE to and disconnect them from a USB port
without shutting down the computer or USB hub. Note that if MAESTRO is running when you
NOTE FOR MULTIPLE USERS ON A NETWORK
There are two ways to reduce the chance that other users will renumber your MCBs:
!
Add the
-I
flag to their MCB Configuration command line, as described in the C
ONNECTIONS
Driver Update Kit instructions. This will allow you to assign whatever ID Numbers you wish,
regardless of the numbers assigned by other users on your network. (Ideally, everyone using
ORTEC instruments on your network should make this change.)
!
To prevent others from renumbering your MCBs (or performing any other actions except read-
only viewing), password-lock your MCBs with the MAESTRO
Lock/Unlock Detector
command. If you lock a detector that will be controlled by a JOB stream, remember to include
the proper password-unlock commands in your
.JOB
file (see the MAESTRO user manual).
8