4.0 After use
Caution – Hot surface
•
Do not touch the well, the insertion tube or the grid
plate - they may be very hot and cause burns.
•
Do not touch the tip of the sensor when it is removed
from the insertion tube/well – it may be very hot and
cause burns.
•
Over 50°C/122°F
If the calibrator has been heated up to temperatures
above 50°C/122°F, you must wait until the instrument
reaches a temperature below 50°C/122°F before you
switch it off.
ETC-125 A only.
Warning……
•
Never leave hot insertion tubes which have been
removed from the calibrator unsupervised – they may
constitute a fire hazard or personal injury.
•
Do not remove the insertion tube from the calibrator
before the insertion tube has cooled down to less than
50°C/122°F.
Caution…
•
The insertion tube must always be removed from the
calibrator after use.
The humidity in the air may cause corrosion oxidation
on the insertion tube inside the instrument. There is a
risk that the insertion tube may get stuck if this is
allowed to happen.
•
If the calibrator is to be transported, the insertion tube
must be removed from the well to avoid damage to the
instrument.
Содержание Jofra ETC-125 A
Страница 1: ...Reference manual Temperature Calibrator Jofra ETC 125 A 400 A 400 R ...
Страница 3: ......
Страница 5: ...2019 04 12 123943 08 List of contents User manual English Reference manual English ...
Страница 7: ......
Страница 8: ...2019 04 12 ON 115 V 1 5 3 4 2 FIG 1 ...
Страница 10: ...2019 04 12 C C AUTO STEP ESC MENU ESC MENU FIG 3 ...
Страница 11: ...2019 04 12 FIG 4 AUTO STEP AUTO STEP ...
Страница 12: ...2019 04 12 FIG 5 ESC MENU ESC MENU ESC MENU ESC MENU ESC MENU ESC MENU ESC MENU ...
Страница 29: ...2 2019 04 12 123943 08 ...