Follow the instructions below before using the calibrator (cf. Fig. 1):
1.
Place the calibrator on an even horizontal surface away from
all draughts.
Caution…
Do not use the instrument if the fan is out of order. Ensure
a free supply of air to the fan (pos. 5).
2.
Check that the voltage shown on the label above the power
control switch is identical to the mains voltage (ETC-400
only). If not, the instrument should be returned to AMETEK
Denmark A/S or the distributor for replacement.
3.
Plug in the cable below (pos. 3) the power control switch
(pos. 4) and check that the earth connection is present.
4.
The selected insertion tube is inserted into the calibrator
(ETC-125 A only).
Caution…
•
The well and the insertion tube must be clean before
use.
•
The insertion tube must never be forced into the well.
The well could be damaged as a result, and the
insertion tube may get stuck.
•
Do not pour any form of liquids into the well. It might
damage the well or cause a hazard.
5.
Place the sensor (pos. 1, Fig. 1).
Содержание Jofra ETC-125 A
Страница 1: ...Reference manual Temperature Calibrator Jofra ETC 125 A 400 A 400 R ...
Страница 3: ......
Страница 5: ...2019 04 12 123943 08 List of contents User manual English Reference manual English ...
Страница 7: ......
Страница 8: ...2019 04 12 ON 115 V 1 5 3 4 2 FIG 1 ...
Страница 10: ...2019 04 12 C C AUTO STEP ESC MENU ESC MENU FIG 3 ...
Страница 11: ...2019 04 12 FIG 4 AUTO STEP AUTO STEP ...
Страница 12: ...2019 04 12 FIG 5 ESC MENU ESC MENU ESC MENU ESC MENU ESC MENU ESC MENU ESC MENU ...
Страница 29: ...2 2019 04 12 123943 08 ...