The following routine must be observed before the insertion tube is
removed and the instrument switched off:
Caution – Hot surface
•
Over 50°C/122°F
If the calibrator has been heated up to temperatures
above 50°C/122°F, you must wait until the instrument
reaches a temperature below 50°C/122°F before you
switch it off.
•
Do not remove the insertion tube from the calibrator
before the insertion tube has cooled down to less than
50°C/122°F
•
Never leave hot insertion tubes which have been removed
from
the
calibrator
unsupervised
–
unauthorized
personnel might pick up insertion tubes causing severe
burns.
•
Never place a hot insertion tube in the optional carrying
case.
Below 0°C/32°F
•
Do not touch the well or insertion tube when these are
below 0°C/32°F - they might create frostbite.
•
If the calibrator has reached a temperature below
0°C/32°F, ice crystals may form on the insertion tube and
on the well. This, in turn, may cause the material
surfaces to oxidize.
To prevent this from happening, the insertion tube and
the well must be dried. This is done by heating up the
calibrator to min. 100°C/212°F until all water left has
evaporated.
It is very important that humidity in the well and insertion
tube is removed to prevent corrosion and frost expansion
damages.
Содержание Jofra ETC-125 A
Страница 1: ...Reference manual Temperature Calibrator Jofra ETC 125 A 400 A 400 R ...
Страница 3: ......
Страница 5: ...2019 04 12 123943 08 List of contents User manual English Reference manual English ...
Страница 7: ......
Страница 8: ...2019 04 12 ON 115 V 1 5 3 4 2 FIG 1 ...
Страница 10: ...2019 04 12 C C AUTO STEP ESC MENU ESC MENU FIG 3 ...
Страница 11: ...2019 04 12 FIG 4 AUTO STEP AUTO STEP ...
Страница 12: ...2019 04 12 FIG 5 ESC MENU ESC MENU ESC MENU ESC MENU ESC MENU ESC MENU ESC MENU ...
Страница 29: ...2 2019 04 12 123943 08 ...