It is very important that humidity in the well and insertion
tube is removed to prevent corrosion and frost expansion
damages.
Caution – ETC-125 A only
The insertion tube must always be removed from the
calibrator after use.
The humidity in the air may cause corrosion oxidation on the
insertion tube inside the instrument. There is a risk that the
insertion tube may get stuck if this is allowed to happen.
Users should/must carry out the following cleaning procedures as
and when required:
•
The exterior of the instrument - Clean using water and a soft
cloth.
The cloth should be wrung out hard to avoid any water
penetrating the calibrator and causing damage.
The keyboard may be cleaned using isopropyl alcohol when
heavily soiled.
•
The insertion tube (ETC-125 A only) - Must always be clean
and should be regularly wiped using a soft, lint-free, dry cloth.
You must ensure there are no textile fibres on the insertion tube
when it is inserted in the well. The fibres may adhere to the well
and damage it.
•
The well - Must always be clean. Dust and textile fibres should
be removed from the well using e.g. compressed air.
Warning
REMEMBER to wear goggles when using compressed air!
•
The target unit surface (ETC-400 R only) - Must always be
clean. Dust and textile fibres should carefully be blown from the
surface with clean air. Do not use fluid to clean the target unit
surface.
Содержание Jofra ETC-125 A
Страница 1: ...Reference manual Temperature Calibrator Jofra ETC 125 A 400 A 400 R ...
Страница 3: ......
Страница 5: ...2019 04 12 123943 08 List of contents User manual English Reference manual English ...
Страница 7: ......
Страница 8: ...2019 04 12 ON 115 V 1 5 3 4 2 FIG 1 ...
Страница 10: ...2019 04 12 C C AUTO STEP ESC MENU ESC MENU FIG 3 ...
Страница 11: ...2019 04 12 FIG 4 AUTO STEP AUTO STEP ...
Страница 12: ...2019 04 12 FIG 5 ESC MENU ESC MENU ESC MENU ESC MENU ESC MENU ESC MENU ESC MENU ...
Страница 29: ...2 2019 04 12 123943 08 ...