32/32
11-06-2001
100975
05
8.
Accept the calculated output scaling factors shown on the screen.
Analog data
Output 1 factor X'
:
4.876
Output 2 factor X'
:
4.876
Output 1 factor Y'
:
585.142
Output 2 factor Y'
:
585.142
Gain in mV/
°
C
'
:
5.000
Offset in
°
C
'
:
0.000
< ESC > aborts
< Enter > to continue
9.
When the calibration is finished the software downloads the new scaling factors for
the analog outputs to the DTI which will take about 1 second.