5.3.5 Resolution
In order to change the resolution on the read-out, you must press the softkey <MENU/Enter>
3 times. The following message will be shown in the display:
RESOLUTION: 0.01
Use the softkey <Reset/SELECT> to scroll between following types of resolution:
RESOLUTION SELECTION MENU
READ-OUT MODE
DTI 100 + 1000
RESOLUTION: 0.01
T1
23.65
°
C
T2
23.60
°
C
│
DTI 1000 only
RESOLUTION: 0.001
T1
23.651
°
C
T2
23.599
°
C
│
DTI 100 + 1000
RESOLUTION: 0.1
T1
23.7
°
C
T2
23.6
°
C
Use the softkey <Menu/ENTER> to select the specific read-out unit.
The sample period is dependent on the selected resolution which can be longer than sample
rate to the display if a resolution above 0.1 is selected. The sample rates for the different
resolutions are shown below:
Resolution
0.1
0.01
0.001
Sample period
1 s
2 s
10 s
Please note that read-out in
Ω
has one more digit than specified in the resolution set-up.