PN 9000-165-VE, Rev P
Flow Calibration
The analyzer must have a flow sensor to configure flow calibration.
Flow sensor calibrations should only be performed on new flow sen-
sors or when the electronics have been changed. Use the same gas
used for O
2
span when calibrating the flow sensor.
Make sure that
Flow
is checked on the
Select Sensor Options
sub-
menu under
Sensor Configuration
.
Figure 2-17.1.
Flow Calibration should only
be done on a new sensor.
1. From the
Main Menu
screen, press
ENTER/MENU
to access the
Calibra-
tion Menu
.
2. Press the Left or Right arrows to locate the
Flow Calibration
menu and
press
ENTER/MENU
.
3. Press
ENTER/MENU
to continue with the calibration or press
CANCEL
to
exit the screen.
4. Turn off the aspirator and calibration gas flow. Press
ENTER/MENU
to con-
tinue or the
CANCEL
to abort the operation.
5. The
Flow Sensor Zero Calibration
screen displays. Press
ENTER/MENU
to
start the zero calibration.
6. The analyzer counts down to zero and when the zero calibration has been
completed, the analyzer will display the
Flow Sensor Span Calibration
screen.
7. You are prompted to inject the Cal gas. Be sure to use aspirator air or zero
gas. Press
ENTER/MENU
to continue and access the
Flow Sensor Span
Calibration
screen.
2-26 | Thermox
®
AMEVision Display User Interface
Содержание AMEVision
Страница 1: ...AMEVision Display User Interface USER MANUAL PN 9000 165 VE Rev P 9000 185 VE Manual Views Rev A ...
Страница 12: ...PN 9000 165 VE Rev P This page intentionally left blank xii Thermox AMEVision Display User Interface ...
Страница 46: ...PN 9000 165 VE Rev P This page intentionally left blank 2 28 Thermox AMEVision Display User Interface ...
Страница 70: ...PN 9000 165 VE Rev P This page intentionally left blank 4 4 Thermox AMEVision Display User Interface ...
Страница 78: ...PN 9000 165 VE Rev P WDG V AMEVision Modbus Map A 2 Thermox AMEVision Display User Interface ...
Страница 79: ...PN 9000 165 VE Rev P EventMask 1 Bit definitions AMEVision Modbus Information A 3 ...
Страница 80: ...PN 9000 165 VE Rev P WDG 5 Command Definitions A 4 Thermox AMEVision Display User Interface ...
Страница 81: ...PN 9000 165 VE Rev P Status Mask Bit definitions AMEVision Modbus Information A 5 ...
Страница 82: ...PN 9000 165 VE Rev P Process Alarm Bit Definitions A 6 Thermox AMEVision Display User Interface ...
Страница 83: ...PN 9000 165 VE Rev P EventMask 2 Bit definitions AMEVision Modbus Information A 7 ...
Страница 84: ...PN 9000 165 VE Rev P This page intentionally left blank A 8 Thermox AMEVision Display User Interface ...