PN 9000-165-VE, Rev P
* The
Namur Standard
is implemented as follows:
4–20 ma Signal
Analyzer Condition
0 mA
Analyzer unpowered, or completely failed
3.5 mA
Critical Alarm – analyzer reading unusable
3.8 mA
Reading Under Range
(Example: User sets range to 2–10 %.)
Current reading is 1.9 %.
4
–
20 mA
Normal Operation
20.5 mA
Reading Over Range
(Example: Range is 0
–
10 %.)
Current reading is 12 %.
1. From the
Main Menu
screen, press
ENTER/MENU
to access the
Configu-
ration Menu
.
2. Press the Left or Right arrows to locate the Analog Output menu and press
ENTER/MENU
.
3. The output that is displayed beneath the sensor name is the active output
you are configuring. Use the arrows to access each of the functions. Press
ENTER/MENU
to view the drop down menu and use the arrow keys to
make your selection.
4. Enter values using the numeric keypad on the AMEVision enclosure. Use
the period for a decimal point.
5. When you have entered the information for each function, navigate to
ACCEPT
and press
ENTER/MENU
after each output has been configured
to accept the values. Press the Left and Right arrows to navigate between
Analog Output 1
,
2
, and
3
.
6. When you have entered the information for each analog output, navigate
to
ACCEPT
and press
ENTER/MENU
after each output has been config-
ured to accept the values.
7. Press
CANCEL
to return to the
Sensor Configuration
menu.
2-12 | Thermox
®
AMEVision Display User Interface
Содержание AMEVision
Страница 1: ...AMEVision Display User Interface USER MANUAL PN 9000 165 VE Rev P 9000 185 VE Manual Views Rev A ...
Страница 12: ...PN 9000 165 VE Rev P This page intentionally left blank xii Thermox AMEVision Display User Interface ...
Страница 46: ...PN 9000 165 VE Rev P This page intentionally left blank 2 28 Thermox AMEVision Display User Interface ...
Страница 70: ...PN 9000 165 VE Rev P This page intentionally left blank 4 4 Thermox AMEVision Display User Interface ...
Страница 78: ...PN 9000 165 VE Rev P WDG V AMEVision Modbus Map A 2 Thermox AMEVision Display User Interface ...
Страница 79: ...PN 9000 165 VE Rev P EventMask 1 Bit definitions AMEVision Modbus Information A 3 ...
Страница 80: ...PN 9000 165 VE Rev P WDG 5 Command Definitions A 4 Thermox AMEVision Display User Interface ...
Страница 81: ...PN 9000 165 VE Rev P Status Mask Bit definitions AMEVision Modbus Information A 5 ...
Страница 82: ...PN 9000 165 VE Rev P Process Alarm Bit Definitions A 6 Thermox AMEVision Display User Interface ...
Страница 83: ...PN 9000 165 VE Rev P EventMask 2 Bit definitions AMEVision Modbus Information A 7 ...
Страница 84: ...PN 9000 165 VE Rev P This page intentionally left blank A 8 Thermox AMEVision Display User Interface ...