PN 9000-165-VE, Rev P
12.
If the sensor is configured for combustibles, inject the Combustibles
Span Gas.
You are then prompted to apply the combustibles span gas to the calibration
gas inlet port of the sensor (tubing must be free of oil and dirt). Press
ENTER/
MENU
when you have applied this calibration gas.
The system will calibrate the combustibles span gas. Press
ENTER/MENU
to accept the gas readings.
13.
If the sensor is configured for methane, inject the Methane Span Gas.
You are then prompted to apply the methane span gas to the calibration
gas inlet port of the sensor (tubing must be free of oil and dirt). Press
ENTER/MENU
when you have applied this calibration gas.
The system will calibrate the methane span gas. Press
ENTER/MENU
to
accept the gas readings.
14. When all gases have calibrated, the screen displays the
Calibration Re-
sults
that include
% drift
as well as recovery time. If the calibration is suc-
cessful, a large GREEN check mark will be displayed. If it is not, a large RED
‘X’ is displayed.
Figure 2-14.2.
Calibration Results screen,
Failed calibration.
Figure 2-14.1.
Calibration Results screen,
Successful calibration.
2-20 | Thermox
®
AMEVision Display User Interface
Содержание AMEVision
Страница 1: ...AMEVision Display User Interface USER MANUAL PN 9000 165 VE Rev P 9000 185 VE Manual Views Rev A ...
Страница 12: ...PN 9000 165 VE Rev P This page intentionally left blank xii Thermox AMEVision Display User Interface ...
Страница 46: ...PN 9000 165 VE Rev P This page intentionally left blank 2 28 Thermox AMEVision Display User Interface ...
Страница 70: ...PN 9000 165 VE Rev P This page intentionally left blank 4 4 Thermox AMEVision Display User Interface ...
Страница 78: ...PN 9000 165 VE Rev P WDG V AMEVision Modbus Map A 2 Thermox AMEVision Display User Interface ...
Страница 79: ...PN 9000 165 VE Rev P EventMask 1 Bit definitions AMEVision Modbus Information A 3 ...
Страница 80: ...PN 9000 165 VE Rev P WDG 5 Command Definitions A 4 Thermox AMEVision Display User Interface ...
Страница 81: ...PN 9000 165 VE Rev P Status Mask Bit definitions AMEVision Modbus Information A 5 ...
Страница 82: ...PN 9000 165 VE Rev P Process Alarm Bit Definitions A 6 Thermox AMEVision Display User Interface ...
Страница 83: ...PN 9000 165 VE Rev P EventMask 2 Bit definitions AMEVision Modbus Information A 7 ...
Страница 84: ...PN 9000 165 VE Rev P This page intentionally left blank A 8 Thermox AMEVision Display User Interface ...