Controller/Interface
|
2-21
PN 9000-165-VE, Rev P
Perform an Automatic Calibration
The functions for automatic calibration will only be visible if you have
an RCU and have configured your sensor for
Automatic Calibration
in the
Sensor Configuration
menu.
There are two ways to perform an automatic calibration:
• Define the automatic calibration time and date in the Auto-Calibration
Configuration (
Auto Cal Config
) menu. This allows the system to perform
calibrations at scheduled times and there is no further user interaction.
• Use the
Start Calibration
(or
Set Cal Gas
) menu to define the gases and
values and then use the
Start Calibration
menu to perform an immedi-
ate calibration. This requires you to program the
Start Calibration
menu
and physically press
ENTER/MENU
to start the calibration. The real-time
sequence of calibration events is displayed on a series of screens which
require user before each step of the calibration continues.
Configuring the Auto Cal Using the Auto-Cal Config Menu
1. From the
Main Menu
screen, press
ENTER/MENU
to access the
Calibra-
tion Menu
.
2. Press the Left or Right arrows to locate the
Auto Cal Config
menu and
press
ENTER/MENU
.
3. Press the Left and Right arrows to navigate to the next screen.
Figure 2-15.
Auto-Calibration
Configuration screen.
Содержание AMEVision
Страница 1: ...AMEVision Display User Interface USER MANUAL PN 9000 165 VE Rev P 9000 185 VE Manual Views Rev A ...
Страница 12: ...PN 9000 165 VE Rev P This page intentionally left blank xii Thermox AMEVision Display User Interface ...
Страница 46: ...PN 9000 165 VE Rev P This page intentionally left blank 2 28 Thermox AMEVision Display User Interface ...
Страница 70: ...PN 9000 165 VE Rev P This page intentionally left blank 4 4 Thermox AMEVision Display User Interface ...
Страница 78: ...PN 9000 165 VE Rev P WDG V AMEVision Modbus Map A 2 Thermox AMEVision Display User Interface ...
Страница 79: ...PN 9000 165 VE Rev P EventMask 1 Bit definitions AMEVision Modbus Information A 3 ...
Страница 80: ...PN 9000 165 VE Rev P WDG 5 Command Definitions A 4 Thermox AMEVision Display User Interface ...
Страница 81: ...PN 9000 165 VE Rev P Status Mask Bit definitions AMEVision Modbus Information A 5 ...
Страница 82: ...PN 9000 165 VE Rev P Process Alarm Bit Definitions A 6 Thermox AMEVision Display User Interface ...
Страница 83: ...PN 9000 165 VE Rev P EventMask 2 Bit definitions AMEVision Modbus Information A 7 ...
Страница 84: ...PN 9000 165 VE Rev P This page intentionally left blank A 8 Thermox AMEVision Display User Interface ...