[SOURce:]VOLT:RANGe[:LEVel]
This command sets the voltage range of the power source. Two voltage ranges are
available: a 135 volt range and a 270 volt range. However, 2003RP units with optional
voltage ranges (2003RP-HV) are available as well. In this case, the actual voltage range
values will be different. To determine the actual range values, use the query format of the
RANGE command. The returned values for high and low range (e.g. 135 and 270) should be
used to change range. In most cases, the high voltage range value will be two times the low
voltage range value. Sending a parameter other than the actual range values will not be
accepted.
For 2003RP-AV models configured for single voltage range, only one range is available. In
this case, the VOLT:RANG commend will return the value for the only available range.
Attempting to select any other voltage range will result in a “–100, Command Error”. The
presence of a single voltage range unit may be determined using the Limit subsystem, see
section 10.7.
Some RP Series may use other voltage range values than shown here. It is recommended
to query the actual range values using the LIM:VOLT? query as part of your application
program. See section 10.7.
When the range is set to 135, the maximum rms voltage that can be programmed is 135
volts.
The VOLT:RANG command is coupled with the CURRent command. This means that the
maximum current limit that can be programmed at a given time depends on the voltage
range setting in which the unit is presently operating.
Command Syntax
[SOURce:]VOLTage:RANGe[:LEVel] <NRf>
Parameters
135 | 270
*RST Value
defined by the PONS
Examples
VOLT:RANG 135
Query Syntax
VOLT:RANGe?
Returned Parameters
<NR2>
Related Commands
VOLT LIM:VOLT? VOLT:RANG:AUTO