6
Calibration
Routine calibration should be performed every 12 months. Non-routine calibration is only
required if a related assembly is replaced or if the periodic calibration is unsuccessful.
Calibration can be performed through the front panel by putting the 2003RP in Calibration
mode (CAL mode) or over the RS232C or IEEE-488 bus if the interface option is installed.
For bus calibration, the PGUI32 program may be used. Refer to the PGUI32 help file
supplied with the program for instructions on calibration over the remote control bus.
All routine calibrations can be performed without removing the top cover (closed case
calibration). For the
–AV option, program 115V or 230V for 135V or 270V respectively.
6.1 Calibration Equipment
Digital Multimeter (DMM):
Two (2) Fluke 8506A or 8508A
100 milliohm Current Shunt:
Isotek Model RUG-Z-R100-0.1 or equivalent
Load Bank:
Various power load resistors will be needed
Model
Range
Current
max.
Load
2003RP
135 V
5.0 A
27
270 V
2.5 A
108
2003RP-AV
115 V
5.9 A
20
230 V
4
2.9 A
78
2003RP-HV
156 V
4.3 A
36
312 V
2.16 A
144
Table 6-1: Load Resistors and Current
4
Some 2003RP-AV may be factory configured for single voltage range only. In this case, the 230 V
range is not available.