2.3 Measurements [Option]
Note: Specifications listed below apply from 300 Hz to 500 Hz. Refer to Supplemental
Specifications for frequencies outside this range.
Parameter
Specification
Unit
True RMS Current
Available on standard unit.
Ranges
0.000 - 4.000
0.00
– 6.00
A
RMS
A
RMS
Accuracy
0.2 % FS + 0.3 % reading
Resolution
0.001
0.01
A
RMS
A
RMS
Peak Current
Ranges
(Remote only) 0.00
– 12.00
0.0
– 20.0
A
PEAK
A
PEAK
Accuracy
0.5 % FS + 0.5 % reading
Resolution
0.01
0.1
A
PEAK
A
PEAK
Crest Factor
Available over bus only. Requires interface option
Range
0.00 - 4.00
Accuracy
0.05
Resolution
0.01
True RMS Voltage
Range
0.0 - 320.0
V
RMS
Accuracy
0.1 % FS + 0.05 % reading
Resolution
0.1
V
RMS
Real Power
Range
0 - 800.0
W
Accuracy
0.2 % FS + 0.5 % reading
Resolution
0.2
W
Apparent Power
Available over bus only. Requires interface option
Range
0 - 800.0
VA
Accuracy
0.3 % FS + 0.8 % reading
Resolution
0.1
VA