Page 12
Signal Processor
User Guide
LMG AE
Fig. 7 - ‘CMD’ (command) input control via a remote switch
234007
Fig. 8 - ‘CMD’ (command) input control via potential-free contacts
234008
Vin
Vin
Remote switch
Potential-free
contacts
CMD
CMD
1
2
1
2
NOTE
Remote switch can be replaced by voltage Vin. Where:
>+3.0V = Switch Open
<+1.5V = Switch Closed
NOTE
Potential-free contacts can be replaced by voltage Vin. Where:
>+3.0V = Open Circuit
<+1.5V = Closed Circuit
Function
Peak Picker
Track & Hold
Valley Picker
Switch Open
Peak Pick
Track
Valley Pick
Switch Closed
Reset
Hold
Reset
Card Type
System 4 Card
All Cards
System 4 Card
Function
Peak Picker
Track & Hold
Valley Picker
Open Circuit
Peak Pick
Track
Valley Pick
Short Circuit
Reset
Hold
Reset
Card Type
System 4 Card
All Cards
System 4 Card