![American TIM-900 Скачать руководство пользователя страница 29](http://html.mh-extra.com/html/american/tim-900/tim-900_installation-and-use-manual_2922166029.webp)
Warning
Press AUTO to confirm the
end of the calibration
dispensing.
In order to get a good calibration
of the TIM-900, use an accurately
Calibrated Container of a capacity
not less than 5 liters
.
Purge all air from the unit
(see
paragraph E7) before carrying out
the calibration.
Carry out the calibration
dispensing at a steady flow-rate
by pulling the trigger completely
and keeping it in open position
until the container is full.
Do not reduce the flow-rate to
reach the graduated area of the
calibrated container.
The correct technique to be used
at the final stages of the filling
operation into the Calibrated
Container is "small topping-ups".
This is achieved by rapidly pulling
the trigger of the TIM-900 and then
releasing it very quickly.
After dispensing, wait a few
minutes to allow the removal of
possible air bubbles from the
Calibrated Container.
Read the Real value only at the
end of this phase as the level in
the container may decrease.
Do not wait more than 15
minutes as the TIM-900 will exit
from the menu and it will no
longer be possible to complete
the calibration operation.
The TIM-900 is ready to accept the
modification of the
indicated value
to
make it correspond to the
real Value
.
Press the "10" key to
increase the indicated
value.
Press the "0.1" key to
decrease the indicated
value.
Warning
Warning
Each time a key is pressed, the
last digit on the right is modified by
one unit.
By holding down the keys, the
value changes, slowly at first
and then rapidly.
E
ng
lis
h
TIM-900
INDICATED
VALUE
Blinking
Calibrated
Container
REAL
VALUE
29