If the TIM-900 is used:
* with fluids having a viscosity close to
the limits of the allowed range (such as
low viscosity antifreeze fluids or high
viscosity oils for gear boxes)
* in extreme flow-rate conditions (close
to the min. and max. values of the
allowed range)
it may be necessary to carry out an
on-site calibration.
The TIM-900 allows the user to carry out
a rapid electronic calibration by
modifying the Calibration Factor (K
Factor).
press key “AUTO”;
it is possible to start the
calibration by
dispensing the fluid into
a calibrated container.
The dispensing operation may be freely
interrupted and resumed.
The Calibration dispensing is finished,
when the level of the fluid reaches the
graduated area of the Calibrated
Container.
During dispensing the
TIM-900 displays:
On-site calibration by
dispensing
into a calibrated container:
The TIM-900 is equipped with a meter
with high-precision oval gears,
pre-calibrated in the factory.
At delivery all TIM-900 are given the
same calibration factor:
K Factor = 1,000
This calibration factor guarantees the
best accuracy in the following
operating conditions:
Fluid: motor oil type 10W 30.
Temperature: 20°C
Flow-rate: 10 liters/min.
The calibration can be done either as:
* an on-site calibration, by dispensing
into a calibrated container, or as
* a direct modification of the calibration
factor.
9.86
Warning
Calibrating
G7
Why calibrate?
How to calibrate
When the TIM-900 displays:
Blinking
Blinking
Blinking
Indicated value
Batch total Dispensed
28