Risk of electric shock!
Use
special care when making measurements, if the
voltages are greater than 25 V~ rms or 35 V .
These voltages are considered a shock hazard.
CAUTION
Risk of injury!
The probe tips are
sharp for accuracy. Be careful when handling and
reattach the probe tip shrouds after use.
• For indoor use. This meter is protected to CAT
lll 600 V, CAT ll 600 V, Pollution Degree 2.
• This product must be used by trained users
only.
• Do not exceed the maximum allowable input
range of any measurement mode.
Mode
Maximum Input
A
~
, A
10 A, 250 V fast
acting fuse (max.
30 seconds every
15 minutes)
mA
~
, mA
200 mA, 250V fast
acting fuse
V
~
, V
600 V
~
rms /
Frequency,
Resistance,
Capacitance, Diode
Test, Continuity
250 V
~
rms /
Surge Protection: 8 kV peak per IEC 61010