![AMALGAMATED INSTRUMENT LD5-RS Скачать руководство пользователя страница 59](http://html1.mh-extra.com/html/amalgamated-instrument/ld5-rs/ld5-rs_operation-and-instruction-manual_2915959059.webp)
5.88
Calculation channel fixed average count
Section:
CC 1
to
CC 8
Display:
A.cnt
Range:
1
to
128
Default Value:
4
Default Access Level
4
Function number
4EF0
to
4EF7
This function set the number of channels to average when the calculation channel is set for fixed average operation
(
F.AUG
).
5.89
Calculation channel multiplication scale
Section:
CC 1
to
CC 8
Display:
Arth SCLE
Range:
Any display value
Default Value:
n/a
Default Access Level
4
Function number
4E90
to
4E97
Arithmetic offset, scale and division functions allow the value sent to the display to be altered prior to display. The
offset operation will be performed first followed by the scale and then the division. The
Arth SCLE
function
allows the user to select a scaling value for the display. For example if this function is set to
2
then the value sent
to the display will (after any addition required) be doubled.
5.90
Calculation channel offset value
Section:
CC 1
to
CC 8
Display:
Arth OFSt
Range:
Any display value
Default Value:
n/a
Default Access Level
4
Function number
4EA0
to
4EA7
Arithmetic offset, scale and division functions allow the value sent to the display to be altered prior to display. The
offset operation will be performed first followed by the scale and then the division. The
Arth OFSt
function
allows the user to select an offset value to be added to the incoming value prior to display. For example if this
function is set to
100
then the value sent to the display will be have the value of 100 added to it.
5.91
Calculation channel division
Section:
CC 1
to
CC 8
Display:
Arth diu
Range:
Any display value
Default Value:
n/a
Default Access Level
4
Function number
4Eb0
to
4Eb7
Arithmetic offset, scale and division functions allow the value sent to the display to be altered prior to display. The
offset operation will be performed first followed by the scale and then the division. The
Arth diu
function
allows the user to select a division value to be used on the incoming value prior to display. For example if this
function is set to
5
then the value sent to the display will (after any addition and scale has been applied) be divided
by 5.
LD5RSMAN-2.0-0