mode. Default settings for this mode are 9600 baud, 7 bit, even parity though this can be altered if the sending unit
is not programmed for these settings. When this mode is selected some of the other input setup functions will not
be seen. An overrange message on the Mettler Toledo display will be displayed as
-OR-
on the PM5 display.
5.35
Ethernet input
Section:
CONF
Display:
E.NEt
Range:
OFF
or
ON
Default Value:
OFF
Default Access Level
CAL
Function number
4d0C
Seen only in LE5 version displays. Sets the Ethernet input to on (
ON
or
OFF
). Note if the Ethernet is being used
to input display data then the RS485 or RS232 inputs cannot be used. Port 10002 must be specified following the
TCP/IP address. When specifying the IP address put a colon ”:” after the IP address followed by the port number
of 10002 e.g. 172.16.243.1:10002. Note this example is not a real IP address.
Modbus TCP can also be used with Ethernet (Function 6 or 16).
5.36
Baud rate
Section:
CONF
Display:
bAud
Range:
300
,
600
,
1200
,
2400
,
4800
,
9600
,
19.2
,
38.4
Default Value:
9600
Default Access Level
CAL
Function number
4d01
Sets the baud rate for the incoming data string. This must be set to match the baud rate from the sending device.
This function will not be seen if Ethernet input is being used.
5.37
Serial input data bits and parity
Section:
CONF
Display:
dAtA
Range:
8.NON
,
8.odd
,
8.Eun
,
7.odd
,
7.Eun
,
7.NON
Default Value:
8.NON
Default Access Level
CAL
Function number
4d02
Sets the number of data and parity bits for the incoming serial string. For example for an incoming serial string
using 7 bit, even parity select
7.EUEN
. This function will not be seen if Ethernet input is being used.
5.38
Select number of channels to scan
Section:
CONF
Display:
CHAN Cnt
Range:
1
,
2
,
3
,
4
,
5
,
6
,
7
,
8
Default Value:
1
Default Access Level
CAL
Function number
4d42
Sets the number of channels to be scanned when set for scanning operation at the
CodE
function.
42 of