MG3 Hot Bar
How to …?
76609576EN-MG3-HB-V1_0a.docx 33/49
CALIBRATION:
Shows the current pressure/force real-values if activated.
For STANDARD calibration adjust resolution and offset.
For 2 point calibration (2P-MODE) proceed as follows:
•
Set RESOLUTION = 1 and OFFSET = 0 and activate CALIBRATION. Then
change to 2P-MODE.
•
Make two measurements with an external load cell: one at low force and
one at high force.
•
The measured voltage value is shown in the display of the MG3. Set this
value in the field INPUT 1 (measuring at low force) resp. INPUT 2
(measuring at high force).
•
The measured force value is shown at the load cell. Set this value in the
field VALUE1 resp. VALUE2.
•
Now deactivate CALIBRATION.
5.7
Set Limit Values
Go to screen
LIMITS
via MENU – LIMITS.
Select up to two indicators to be monitored per channel.
Notes
•
The limit value monitoring is analyzed after each measurement.
The evaluation result is:
shown by the LEDs on the front side.
logged.
available at the digital output on the rear side.
If both channels are monitored and the measurements are
simultaneously, the evaluation result is formed by a logical disjunction.
Defining
limits
MAX LIMIT, MAX WARNING, MIN WARNING, MIN LIMIT
Set the minimum and maximum prewarn and limit values.
The limit value monitoring is deactivated if the min and max values are set to
zero. The same applies for the prewarn values.