22
Language: English Model: Netti III HD Version: February 2013
Height of foot plate:
The foot plates are stepless height adjustable.
• Unfix the adjustment screw so that the
adjustment bar moves freely.
• Adjust the foot plate to required height,
then tighten the screw.
Angle of foot plate:
• Unfix the screw as shown below using an
Allen key.
• Adjust the foot plate to the required angle
and tighten the screw.
5 mm Allen key.
As accessory an adjustment wheel can
be used instead of a screw.
Locking and releasing the foot plates
• The foot plates come with a locking bolt
which makes the plates stronger
• To lock the foot plates pull the plastic
lock on the right foot plate and
place the lock over the bolt on the left
foot plate.
• To release the foot plate pull the plastic
lock and lift the right foot plate up.
While making the adjustment there
must be no load on the foot plates.
For outdoor use, there should be a
clearance of 4-5 cm between the
foot plate and the ground.
Never stand on the foot plates due
to the risk of tipping forward.
When adjusting foot support
in angle, be aware of squeeze hazard
between moving parts.
Removing the foot support:
• Pull the plastic lock on the foot plate
rearwards, so that the pin is released,
and the foot plate can be folded up.
• Raise the foot support a few degrees.
• Release the foot support by pulling it
slightly straight up.
• Swing the foot support outwards.
• Lift and remove the foot support. (see
illustration next page).