background image

26

www.humancaregroup.com

 Functions

Emergency stop

In an emergency, the emergency stop can be 

activated by pressing the RED button. To reset 

the emergency stop, turn the RED button 

clockwise.

Emergency lowering

In event of an emergency the emergency low-

ering function can be used. Press emergency 

lowering button and down button at least 3 

seconds simultaneously to activate the emer-

gency lowering. Older lifters may not have 

double key command and delay.

An audio signal indicates that the lift is lowering. 

Always ensure that emergency lowering can be 

performed as safely as possible for the patient. 

The emergency lowering button only works 

when the lift is switched on and emergency stop 

has been deactivated. When the emergency 

lowering is active, the end position switches are 

inactive. If the button is pushed until the lift is 

in its lowest position, the lift will start to go up 

again. Always stop at the lowest position.

o

 The emergency lowering button 

must only be used in emergency 

situations.

Safety catch

The lift is equipped with a safety catch. This 

prevents involuntary falls. If the lift belt is 

pulled out too quickly, the safety catch will 

engage, thus preventing the fall. Slacken the lift 

belt to unlock the catch. With the safety catch 

engaged, the lever (10) can never be turned, as 

this blocks the lowering function of the lift!

Mechanical emergency lowering

During power loss and the hoist is loaded, user in 

the hoist, the mechanical emergency lowering is 

activated and the hoist is lowering in a safe and 

reliable manner.

Overheating

The lift has overheating protection which stops 

the motor if it gets too hot. Overheating can 

occur if the motor is overloaded or operated 

continuously for quite a long time. The lift’s 

work cycle (work:pause) is 15:85. Maximum 

continuous operation is two minutes. If this 

is exceeded, there is a risk of the motor 

overheating.

 Symbols on the unit and product label

f

CE marking

g

Date of manufacture

h

Manufacturer

j

Max. load

 Instructions: Read the instructions 

and make sure you understand them 

before using this product

l

Article number/reference number

n

Serial number

ENG

ENG

o

Warning

q

 The lift contains lead batteries and 

electronics and it must be recycled 

appropriately
Third party certification marking  

      (not applicable for all regions and 

versions)

p

Type-B product

Switch the product off/Switch the 

product on

Содержание 55030

Страница 1: ...Altair Anv ndarmanual User manual Manuel de l utilisateur Gebruikshandleiding Benutzerhandbuch...

Страница 2: ...raise and lower patients in a vertical direction and never diagonally or through angles The lift must always be positioned directly above the patient when lifting The patient must always be positioned...

Страница 3: ...th Altair Before use always ensure that personnel who use the equipment have been given adequate tuition and training in use of the lift and its accessories the lift belt is not twisted or worn and th...

Страница 4: ...hanical Class Internally powered equipment Type B IP rating lift IPX4 IP rating hand control IPX5 Noise level 65dB loaded and unloaded Lifting speed 1 8 m min 3 0 m min unloaded only Lifting height 20...

Страница 5: ...www humancaregroup com Product description 1 Hand control 2 Lift belt 3 Telescopic bar 4 Leg hook 5 Secondary control panel 6 Emergency lowering button 7 Indicator 8 Emergency stop ENG 1 2 3 8 7 6 5...

Страница 6: ...duct has been tested by an accredited test institute and complies with the necessary requirements in accordance with the Medical Devices Directive for Class 1 products MDD 93 42 EEC The lift is CE mar...

Страница 7: ...falls If the lift belt is pulled out too quickly the safety catch will engage thus preventing the fall Slacken the lift belt to unlock the catch With the safety catch engaged the lever 10 can never be...

Страница 8: ...run at two speeds If the Up or Down button is pressed for longer than 3 seconds the lift will run in the faster mode If the lift is loaded with more than approx 40 kg it will only run in the slower m...

Страница 9: ...USB for communicating with the lift Secondary control panel The lift is equipped with a secondary control panel located on the lift which can be used as an alternative to the hand control It is equipp...

Страница 10: ...level is Low The lift must be charged up immediately Charger The indicator on the lift flashes red when the batteries need recharging Plug the charger cable into the hand control or put the hand cont...

Страница 11: ...patient Press the Down button on the hand control or control panel to lower the patient Pressing the button activates a soft start and releasing it causes a soft stop Lower the lift so that it stands...

Страница 12: ...ve three adjustable positions To change position turn the hooks 1 4 turn while pushing or pulling the hooks to the desired position The hooks will automatically fall into place in the position and so...

Страница 13: ...atient individually and in accordance with their needs Consult the user manual for the relevant sling to make sure the product is used correctly Human Care recommends using only Human Care slings with...

Страница 14: ...or the Down button on the hand control or the control panel Constant yellow Charging in progress Interrupt charging Nothing happens when you press the Up button on the hand control or the control pane...

Страница 15: ...he lift belts for damage or wear and tear Feed out enough of the lift belts to allow Care instructions for the lift and accessories their entire length to be checked Contact a retailer if the belt is...

Страница 16: ...vel Compliance level Electro magnetic environment guidance Electrostatic discharge ESD IEC 61000 4 2 6 kV contact 8 kV air 6 kV contact 8 kV air Floors should be wood concrete or ceramic tile If floor...

Страница 17: ...4 3 3 Vrms 150 kHz to 80 MHz 3 V m 80MHz to 2 5GHz 10 V m 800MHz to 2 5GHz 3 Vrms 3 V m 10V m Portable and mobile RF communications equipment should be used no closer to any part of the Human Care li...

Страница 18: ...e lifts The Human Care lifts are intended for use in an electromagnetic environment in which radiated RF disturbances are controlled The customer or the user of the Human Care lifts can help prevent e...

Отзывы: