background image

www.humancaregroup.com

33

 Troubleshooting

Description

Indicator

Action

Nothing happens when you press the Up 

button or the Down button on the hand 

control or the control panel

Not lit

1) Check that the mains switch is on.
2) The batteries are fully discharged. Charge the lift 

immediately
3) The emergency stop is pushed in. Deactivate by turning 

the red button clockwise.

Nothing happens when you press the Up 

button or the Down button on the hand 

control

Flashing green
Flashing yellow
Flashing red

Check the hand control. If the hand control is defective it 

needs to be replaced. Contact a service technician.

Nothing happens when you press the Up 

button or the Down button on the hand 

control or the control panel.

Constant red

1) Lift motor is overheated. Wait a while and try again.
2) Lift is overloaded. Lower using the emergency lowering 

button

Nothing happens when you press the Up 

button or the Down button on the hand 

control or the control panel.

Constant yellow

Charging in progress. Interrupt charging.

Nothing happens when you press the Up 

button on the hand control or the control 

panel.

Flashing green
Flashing yellow
Flashing red

The lift is in its maximum upper position. Press the Down 

button on the hand control or the control panel to lower 

the lift.

Nothing happens when you press the 

Down button on the hand control or the 

control panel.

Flashing green
Flashing yellow
Flashing red

The lift is in its maximum lower position. Press the Up 

button on the hand control or the control panel to raise 

the lift.

The belt is not fed out when the lift is on 

a surface.

Flashing green
Constant green
Flashing yellow
Flashing red

The lift belt has to be pulled upwards at the same time as 

the Down button on the hand control or the control panel 

is held pressed in.

The indicator switches from flashing 

green or yellow to flashing red.

Flashing red

The batteries need charging but you can still lift the 

patient a few times. The lift should be set to charge as 

soon as possible.

ENG

Содержание 55030

Страница 1: ...Altair Anv ndarmanual User manual Manuel de l utilisateur Gebruikshandleiding Benutzerhandbuch...

Страница 2: ...raise and lower patients in a vertical direction and never diagonally or through angles The lift must always be positioned directly above the patient when lifting The patient must always be positioned...

Страница 3: ...th Altair Before use always ensure that personnel who use the equipment have been given adequate tuition and training in use of the lift and its accessories the lift belt is not twisted or worn and th...

Страница 4: ...hanical Class Internally powered equipment Type B IP rating lift IPX4 IP rating hand control IPX5 Noise level 65dB loaded and unloaded Lifting speed 1 8 m min 3 0 m min unloaded only Lifting height 20...

Страница 5: ...www humancaregroup com Product description 1 Hand control 2 Lift belt 3 Telescopic bar 4 Leg hook 5 Secondary control panel 6 Emergency lowering button 7 Indicator 8 Emergency stop ENG 1 2 3 8 7 6 5...

Страница 6: ...duct has been tested by an accredited test institute and complies with the necessary requirements in accordance with the Medical Devices Directive for Class 1 products MDD 93 42 EEC The lift is CE mar...

Страница 7: ...falls If the lift belt is pulled out too quickly the safety catch will engage thus preventing the fall Slacken the lift belt to unlock the catch With the safety catch engaged the lever 10 can never be...

Страница 8: ...run at two speeds If the Up or Down button is pressed for longer than 3 seconds the lift will run in the faster mode If the lift is loaded with more than approx 40 kg it will only run in the slower m...

Страница 9: ...USB for communicating with the lift Secondary control panel The lift is equipped with a secondary control panel located on the lift which can be used as an alternative to the hand control It is equipp...

Страница 10: ...level is Low The lift must be charged up immediately Charger The indicator on the lift flashes red when the batteries need recharging Plug the charger cable into the hand control or put the hand cont...

Страница 11: ...patient Press the Down button on the hand control or control panel to lower the patient Pressing the button activates a soft start and releasing it causes a soft stop Lower the lift so that it stands...

Страница 12: ...ve three adjustable positions To change position turn the hooks 1 4 turn while pushing or pulling the hooks to the desired position The hooks will automatically fall into place in the position and so...

Страница 13: ...atient individually and in accordance with their needs Consult the user manual for the relevant sling to make sure the product is used correctly Human Care recommends using only Human Care slings with...

Страница 14: ...or the Down button on the hand control or the control panel Constant yellow Charging in progress Interrupt charging Nothing happens when you press the Up button on the hand control or the control pane...

Страница 15: ...he lift belts for damage or wear and tear Feed out enough of the lift belts to allow Care instructions for the lift and accessories their entire length to be checked Contact a retailer if the belt is...

Страница 16: ...vel Compliance level Electro magnetic environment guidance Electrostatic discharge ESD IEC 61000 4 2 6 kV contact 8 kV air 6 kV contact 8 kV air Floors should be wood concrete or ceramic tile If floor...

Страница 17: ...4 3 3 Vrms 150 kHz to 80 MHz 3 V m 80MHz to 2 5GHz 10 V m 800MHz to 2 5GHz 3 Vrms 3 V m 10V m Portable and mobile RF communications equipment should be used no closer to any part of the Human Care li...

Страница 18: ...e lifts The Human Care lifts are intended for use in an electromagnetic environment in which radiated RF disturbances are controlled The customer or the user of the Human Care lifts can help prevent e...

Отзывы: