The internal battery is a 3.7V, 750mAh Lithium-Polymer encapsulated rechargeable cell. The
Icon indicates the battery voltage as follows: -
OFF
The Icon fills from the left when the external power supply is connected. This indicates that the
battery is receiving a charging current and should be fully charged within 14 hours.
Note: It is recommended that the unit be charged overnight when first received to ensure the battery is
fully charged.
Note: The 4-20mA output should not be operated while the instrument is charging.
14
Performing an AutoCal
Before an AutoCal can be performed successfully, the user must have a reference moisture
reading. This reading is entered into the SADPminiEx in the AutoCal process as the AutoCal
value. The AutoCal routine can be performed using a known reference gas or under ambient
conditions.
Note: The AutoCal process will automatically correct any SPCs (see section 9.2.5) already in the SADPminiEx.
Less than 3.3V dc
3.3 to 3.38V dc
3.38 to 3.46V dc
3.46 to 3.54V dc
3.54 to 3.62V dc
Greater than 3.62V dc