Page 23 of 33
11.2.9
Unit Address
Description – It is possible to communicate with up to 32 SADPminiEx units via the
RS485. However, for this to operate correctly each unit must have a different address.
The default address setting for the SADPminiEx is 1.
Operation – In the ‘Logging Menu’ select the ‘Unit Address’ option. The user is then
required to enter the required address using the standard numerical editor format.
Pressing the ‘ ’ key confirms the new address.
11.2.10
Free Space
Description – This routine allows the user to see how much space is available in the
SADPminiEx memory, in terms of logged readings.
Operation – In the ‘Logging Menu’ select the ‘Space Free’ option. The screen
automatically reverts to the ‘Logging Menu’ after 3 seconds.
11.2.11
Exit
Description – The Exit option allows the user to return to the normal display.
Operation – In the ‘Logging Menu’ select the ‘Exit’ option.
12
Pressure Correction
When a line pressure is entered (see 9.1.2) and ºC or ºF have been selected the displayed
moisture value is pressure corrected. The fact that pressure correction is active is indicated
by the phrase “@ 10 barg” (or some variation) been displayed.
Pressure corrected values are also used in the External Log and Internal Log screens when
ºC or ºF are the selected units.
Warning: This function is purely an internal calculation and DOES NOT
allow an elevated pressure to be applied to the measuring head.
13
Battery Current Status
Only use the Charger supplied with SADPminiEx
The battery Icon in the bottom left-hand corner of the display indicates the status of the battery.