9.2 Measuring Point Designation
Each measuring point can be given a 10-digit alphanumeric designation to
optimally identify the type of sensor, the measuring location or the purpose of
the application. This comment will be indicated with all standard displays of
measured values. If it has not been programmed, the abbreviation of the
measuring range will be indicated. In case of outputs via interface the
measuring point designation appears in the program header as ´COMMENT´
and in the list of measured values (see man. 6.6.1).
Input in function ´Comment´, see 5.5
Comment: Temperature
9.3 Averaging Mode
The types of averaging, which are defined through the function ´Averaging
Mode´, are described in section 6.4.2.
Function ´Averaging Mode´ no averaging:
Averag.Mode: -----
9.4 Locking the Programming of the Sensor
see man. 6.3.12
The function parameters of each measuring point are protected by the locking
mode up to an adjustable locking level. Before any programming is performed
the locking mode must be correspondingly lowered. If a dot is indicated
following the locking mode on the display then a modification is not possible.
Locking Level
Locked Functions
0
none
1
Meas. range + element flags
3
+ Dimension
4
+ Zero point and gain (slope) correction
5
+ Base value, factor, exponent
6
+ Analog output, start and end
7
+ Limit values, max and min
Function ´LockingMode´:
Locking: 5
In the menu ´Sensor Programming´ the functions are arranged from top to
bottom so that the locked functions cannot be selected. The menu and keypad
locking must be used to protect the programming and the process control
during a measurement against unauthorised modification (see 5.6).
ALMEMO
®
5990-2
53
Sensor Programming