Cancelling the memory output using the key:
ESC
Interrupting the memory output using the key:
<
STOP
>
The following options are available after an output interruption:
Recall individual measured values:
<
MANU
>
Re-start the automatic output:
<
START
>
Cancelling the memory output using the key:
ESC
After the memory output the device returns to the menu ´Memory´.
The memory contents will be output with the same print format as with a printer
operation, including multiple printouts and different formats (see manual 6.6.1).
Print Format:
MEMORY: 03
(evtl. connector number)
NUMBER: 12-001
(if activated)
DATE: 12.03.00
List format
12:30:00 01: +0012.0 °C NiCr Designation
below each other
02:!+0008.8 °C NiCr Water
03:>+125.00 °C Ntc MotorOil
The connector number of an external memory will be printed after
each headline ´MEMORY´.
For reading Smart-Media-Cards an own reader ZA1904SLG is avail-
able. Commercial reader aren't usable on reason of the memory for-
mat!
Clear Memory
Menu ´Memory´:
Select function ´MemoryFree´ (see 5.4):
No:-- Free:
kB
0217.
To clear the memory, first press the key:
PROG
the memory capacity is highlighted:
No:-- Free:
kB
0217.5
then the key:
<
CMEM
>
Full mem. capacity is indicated as avail. mem. space:
No:--Free:0507.5kB
Or cancel by using the key:
ESC
50