FORM DOC 120 A
MU 7093 EN C
DUAL TRONIQUE
Page 51/69
This document is available on www.alma-alma.fr
LF COEFFICIENT (K1)
: Coefficient for low flow. The unit depends on settings (pulses/liter or
pulses/kg)
LOW FLOW/K1 (Q1)
: Reference low flow so that [flowmin]≤Q1<[flowminx1.5]. According to
the flow unit
HF COEFFICIENT (K2)
: Coefficient for high flow. The unit depends on settings (pulses/liter
or pulses/kg)
HIGH FLOW/K2 (Q2)
: Reference high flow so that [flowminx3]≤Q2<[flowmax]. According to
the flow unit
METER COEFFICIENT
0
0.0000
COEF FICIENT K1 EMA
00.0000
LF COEFFICIENT (K1)
0
00.0
m
3
LOW FLOW/K1 (Q1) /h
0
0.0000
COEF FICIENT K2 EMA
00.0000
HF COEFFICIENT (K2)
0.000
m
3
HIGH FLOW/K2 (Q2) /h
0
00.0
m
3
HIGH FLOW/K2 (Q2) /h
0.000
m
3
LOW FLOW/K1 (Q1) /h
7.3.2 Sub-menu VISCOSITY CORRECTION
This menu is used to define the correction to be applied to the low viscosity product when
it is defined with correction (SUPERVISOR mode). See the marking of the meter or the
calibration certificate. Maximum value:
0.4%.
+0.0
VISCOSITY CORRECTION
+
0.0
VISCOSITY CORRECTION
7.3.3 Sub-menu UNIT
Choose the accuracy of the quantity and the unit of the flow that will be displayed and
printed for the EMA measuring system.
UNIT
ACCURACY
FLOWRATE
ACCURACY
: Choose the accuracy of the quantity that will be displayed and printed.
According to the unit set in menu
CONFIGURATION>UNIT>QUANTITY
(measure of a volume
or a mass).
123456
ACCURACY
1234.56
ACCURACY
12345.6
ACCURACY
ACCURACY
Unit = L or kg
Selon CONFIGURATION>UNIT>QUANTITY
FLOWRATE
: Choose the accuracy of the quantity that will be displayed and printed.
According to the unit set in menu
CONFIGURATION>UNIT>QUANTITY
(measure of a volume
or a mass).