CONFIGURATION>UNIT>QUANTITY
L
FLOWRATE
123.4
m
3
FLOWRATE /h
1234
L
FLOWRATE /min
12345
L
FLOWRATE /h
CONFIGURATION>UNIT>QUANTITY
KG
FLOWRATE
12345
Kg
FLOWRATE /h
1234
Kg
FLOWRATE /min
123.4
FLOWRATE /h
tons/h, (
there’s no pictogram for tons)
7.3.4 Sub-menu METER FLOWRATES
The accuracy and the unit of the displayed values are specific to the measuring system
and depend on the choices made during the metrological configuration
EM>UNIT menu
.
MINIMUM FLOWRATE:
Set the metrological minimum flowrate of the EMA measuring
system
MAXIMUM FLOWRATE:
Set the metrological maximum flowrate of the EMA measuring
system
METER FLOWRATES
0
04.0
m
3
ENTER MINIMUM FLOW /h
4.0
m
3
MINIMUM FLOWRATE /h
0
50.0
m
3
ENTER MAXIMUM FLOW /h
50.0
m
3
MAXIMUM FLOWRATE /h
7.3.5 Sub-menu QUANTITIES
The accuracy and the unit of the displayed values are specific to the measuring system
and depend on the choices made during the metrological configuration
EM>UNIT menu
.
200
L
MINIMUM QUANTITY
00
0
200
MINIMUM QUANTITY
00
L
MANIFOLD QUANTIT Y
0
0
L
MANIFOLD QUANTIT Y
QUANTITIES
5
L
FIXED QUANTITY
00
L
RELEASE QUANTITY
Not applicable
Not applicable
MINIMUM QUANTITY
:
Set the minimum quantity of the EMA measuring system. This value
is given by the association of the turbine meter, the MIC and other parts of the
measuring system.