WARNING:
Failure to use correct lifting methods could cause patient fall, injury, or
equipment damage.
6.
With the help of another person, do as follows:
a.
Align the ends of the support top with the pre-determined hole
location. If you are to install a Flat Top without a pad installed,
make sure the side with the Velcro® tape is up (patient side).
NOTE:
You may have to adjust the length of the Table so that the columns are at
the correct distance apart to fit the support top “Adjust the Length of the
Table” on page 19.
WARNING:
Never remove a cross pin from a support top while a patient is on the
support top. Patient fall or injury could occur.
b.
At each end of the support top,
insert a cross pin into the
H-bracket, through the support
top connector assembly, and
through the other end of the H-
bracket.
WARNING:
Do not use the support top without the
cross pins installed correctly through the H-brackets. To do so could cause
patient fall, injury, or equipment damage.
c.
Make sure the drop lock on the end of the cross pin moves to the
side to hold the support top in position.
Remove a Support Top and/or Bracket
WARNING:
Never remove a cross pin from a support top while a patient is on the
support top. Patient fall or injury could occur.
1.
Make sure the patient has been moved from the support top.
2.
Do the applicable steps of the install procedure in reverse order.