Allen® Advance Table User Manual (D-720686-A6)
Install the Lateral Top
NOTE:
You may have to adjust the length of the Table so that the columns are at
the correct distance apart to fit the support top. Refer to “Adjust the
Length of the Table” on page 19.
NOTE:
The lateral top will only install on the lateral bracket. If the H-brackets or a
Advance Pinless H-Bracket are installed on the table, go to “Install the
Lateral Bracket” on page 28.
WARNING:
Make sure that you put the patient’s head at the head end of the lateral
top. The foot end of the lateral top is not designed to support the weight
of the patient’s torso. Personal injury, patient fall, and/or equipment
damage can occur.
1.
Position the head end of the lateral top at the head end of the Allen®
Advance Table.
•
The head end of the Allen® Advance Table is the end with the
control panel.
•
The head end of the lateral top is the end that does not articulate,
and has the air manifold.
2.
At each end of the lateral top, install a cross
pin into the lateral bracket, through the
lateral top connector assembly, and the
other side of the lateral bracket. Make sure
that the cross pins that secure the lateral
top are in the same orientation as the cross
pins that secure the lateral bracket.
WARNING:
Do not use the lateral top without the cross pins installed correctly
through the lateral bracket. To do so could cause patient fall, injury, or
equipment damage.
3.
Make sure the drop lock on the end of the cross pin moves to the side
to hold the support top in position.
4.
Connect the lateral top power cord to the lateral top power port at
the foot end of the Allen® Advance Table base unit.
5.
Install the lateral top pad on the lateral top.