Allen® Advance Table User Manual (D-720686-A6)
WARNING:
The maximum weight load for the traction assembly is 50 lb (22.6 kg). To
exceed this weight could cause patient fall, injury, or equipment damage.
WARNING:
Make sure the traction weight is in a safe position, free from possible
unintended movement. Failure to do so could cause patient fall, injury, or
equipment damage.
WARNING:
Do not press the Cervical Traction Release button while traction weight is
being applied. To do so could cause the assembly to retract and the
traction weight to move. Patient fall, injury, or equipment damage could
occur.
6.
Use traction weight per facility protocols. Make sure that—
•
The traction weight is in a safe position such that it is free from
possible unintended movement.
•
The traction rope stays in the indented section of the cervical
traction bar.
•
You do not press the
Cervical Traction
Release
button while
traction weight is being applied.
NOTE:
The cervical traction bar is an optional accessory that is supplied to help
with the angle of pull for the cervical traction. If used, install the bar at the
applicable location on the H-bracket.
7.
After use of the traction weight, remove the weight and the rope from
the assembly.
8.
Press the
Cervical Traction
Release
button, and then push the
assembly into the column so that it is in the locked-in position.