Page 13
The PTT provides “EZ-Plug & Play” with Mountz Transducers that feature “ARCII” technology, an instant auto-
recognition system of the transducer connected to the PTT. When an ARCII Transducer is connected to the PTT it
automatically recognizes the transducer and displays the Model and Serial Number of the connected transducer on
the PTT.
The information stored in the ARCII chips contains:
The Model of Transducer
The Serial Number of the Transducer
True Calibration Information
Date of Calibration
Auto Clear
Angle Threshold
T+A Threshold
First Peak
Direction
Clock
Clock Setting
2:57 PM
11/15/05
Main
1. Press the
"Menu"
Button.
2. Use the
Scroll Down
key and select Setup by highlighting and pressing
Enter
.
3. Use the
Scroll Up or Down
key to toggle through the Setup options.
4. Press
Enter
key to select a setup setting for
Clock
.
Filter
Units
Mode
Setup
Selecting Clock
1. Use the
Scroll Up or Down
key to toggle through the digits for the date
and time.*
2. Use the
Right and Left
key to move through the time and date sections.
3. Press
Enter
key once setting is complete.
4. Press
Main
button to return to main display screen.
Note:
Must toggle through the first “time digits” to change from AM to PM.
Six months from the date of the transducer’s calibration a message will appear on the screen informing the
operator it has been six months since the date of calibration. At this point the operator can pull the Transducer out of
service or decide to continue to use it. After the initial message the user will be reminded one a month.
T r a n s d u c e r C a l i b r a t i o n R e m i n d e r
A R C I I ( A u t o R e c o g n i t i o n C h i p )