background image

Tolerance
Auto Clear
Angle Threshold
T+A Threshold
First Peak
Direction

Auto Clear Setting

Mode: Auto
Time (seconds): 2

Main

1. Press the 

"Menu"

Button. 

2. Use the 

Scroll Down

key and select Setup by highlighting and pressing 

Enter

.

3. Use the 

Scroll Up or Down

key to toggle through the Setup options.

4. Press 

Enter 

key to select a setup setting for 

Auto Clear

.

Filter

Units

Mode

Setup

Selecting Clear

1. Use the 

Scroll Up or Down

key to toggle between Manual or Auto Clear.

2. Press the

Scroll Left or Right 

key to move down to the Time selection

3. For Auto Clear, use the 

Scroll Up or Down

key to toggle between the selection

of time between 1- 5 seconds.

4. Press 

Enter 

key once setting is complete.

5. Press 

Main 

button to return to main display screen.

Tolerance
Auto Clear
Angle Threshold
T+A Threshold
First Peak
Direction

This is for the Audit mode. The default setting is 2 degrees, but it can be set from 1 to 5 degrees.

Angle Threshold

1. Press the 

"Menu"

Button. 

2. Use the 

Scroll Down

key and select Setup by highlighting and pressing 

Enter

.

3. Use the 

Scroll Up or Down

key to toggle through the Setup options.

4. Press 

Enter 

key to select a setup setting for 

Angle Threshold

.

Filter

Units

Mode

Setup

Selecting Angle Threshold

Page 9

Содержание PTT 2000

Страница 1: ...Mountz Inc ISO 9001 2000 company P T T P T T Portable Torque Tester Operating Instructions 2000 ...

Страница 2: ... Hz Easy to use Menu Structure Real Time Clock for time stamping of readings Six digit display USB interface to download readings to PC High Capacity Li Ion Batteries for long life 30 hrs with standard transducers and 16 hrs with brushless rotary Can connect to most mv v transducers and can store calibration data for up to 50 non smart torque sensors The 5VDC capability allows unit to be used with...

Страница 3: ...Signal Negative output from bridge transducer 8 Not Used 9 Angle Lead TTL output from angle detector in angle transducer 10 Sensor Drive Used to detect transducers 11 Bridge Signal Positive output from bridge transducer 12 Shield 13 Angle Trail TTL output for angle lags Angle Lead by 90 degrees 14 5 Volt Angle Supply 15 Data for Smart Transducer Proprietary ARCII protocol for smart transducers fro...

Страница 4: ...ter Units Mode Setup Main Menu 3000 Hz 2000 Hz 1500 Hz 500 Hz 200 Hz ozf in lbf in lbf ft kgf cm kgf m N m cN m Track Peak First Peak Audit Torque Angle Tolerance Auto Clear Angle Threshold T A Threshold First Peak Direction Clock Sub Menu Sub Menu Sub Menu Sub Menu Enter Enter Enter Enter Contrast ...

Страница 5: ...erial Number 04 93 071 If an un calibrated transducer is connected then the Analyzer will display a screen as below Note The PTT supports Non Smart transducers and Non Mountz transducers Non smart transducers can be detected and can be calibrated in the internal memory The PTT will offer 2 choices It offers to calibrate or choose from a list of transducers stored in the internal memory Mountz PTT ...

Страница 6: ... or Down key to toggle through Filter Units Mode and Setup 2 Press Enter key to select a choice Filter Units Mode Setup 1 Select Filter by highlighting and pressing Enter 2 Use the Scroll Up or Down key to toggle through the Filter options 3 Press Enter key to select a Filter Setting Filter Units Mode Setup 3000 Hz 2000 Hz 1500 Hz 500 Hz 200 Hz Selecting Filters Main Main Cancel ...

Страница 7: ...pressing Enter 2 Use the Scroll Up or Down key to toggle through the Mode options 3 Press Enter key to select a Mode Setting 4 The Main or reading screen will display the value of the selected Mode Filter Units Mode Setup Selecting Mode Options Tolerance Auto Clear Angle Threshold T A Threshold First Peak Direction 1 Select Setup by highlighting and pressing Enter 2 Use the Scroll Up or Down key t...

Страница 8: ...pe mechanical torque wrenches The PTT captures the point where the wrench clicks This peak may be used for operator training on correct use of the wrench Always apply torque smoothly to avoid false first peak readings See page 11 for setting up First Peak First Peak FPK 3000 Hz lbf in Menu Tool Tests FPK 3000 Hz 5 4373 lbf in Menu Tool Tests 7 6261 This mode is used to determine first movement or ...

Страница 9: ...ggle through the Setup options 4 Press Enter key to select a setup setting for Tolerance Filter Units Mode Setup Selecting Tolerance 1 Press the Right or Left key to toggle between High and Low Tolerance 2 Use the Scroll Up or Down key to change the tolerance settings 3 Press Enter key once setting is complete 4 Press Main button to return to main display screen This function controls the method o...

Страница 10: ... down to the Time selection 3 For Auto Clear use the Scroll Up or Down key to toggle between the selection of time between 1 5 seconds 4 Press Enter key once setting is complete 5 Press Main button to return to main display screen Tolerance Auto Clear Angle Threshold T A Threshold First Peak Direction This is for the Audit mode The default setting is 2 degrees but it can be set from 1 to 5 degrees...

Страница 11: ...Main 1 Press the Menu Button 2 Use the Scroll Down key and select Setup by highlighting and pressing Enter 3 Use the Scroll Up or Down key to toggle through the Setup options 4 Press Enter key to select a setup setting for Torque Angle Filter Units Mode Setup Selecting Torque Angle 1 Press the Right or Left key to toggle between the three settings Angle Torque Allow Ratchet 2 Use the Scroll Up or ...

Страница 12: ...hrough the sensitivity settings Low Medium High 2 Press the Right or Left key to move down to the Min Peak setting location 3 Use the Scroll Up or Down to set the Minimum Peak 4 Press Enter key once setting is complete 5 Press Main button to return to main display screen Page 11 Note First Peak Sensitivity This feature is provided for click wrenches It is not relevant for Mountz Break over or Cam ...

Страница 13: ...irection Setting Direction BOTH Main 1 Press the Menu Button 2 Use the Scroll Down key and select Setup by highlighting and pressing Enter 3 Use the Scroll Up or Down key to toggle through the Setup options 4 Press Enter key to select a setup setting for Direction Filter Units Mode Setup Selecting Direction 1 Use the Scroll Up or Down key to toggle through the dIrection settings Both CW CCW 2 Pres...

Страница 14: ...he Scroll Up or Down key to toggle through the Setup options 4 Press Enter key to select a setup setting for Clock Filter Units Mode Setup Selecting Clock 1 Use the Scroll Up or Down key to toggle through the digits for the date and time 2 Use the Right and Left key to move through the time and date sections 3 Press Enter key once setting is complete 4 Press Main button to return to main display s...

Страница 15: ... unit Just press a key and the unit will power on When the unit has not been used for a designated time frame it will enter a Sleep Mode The unit features a processor that checks to see if a key pressed while it is in the Sleep Mode There is no significant drain on the batteries The PTT is powered using the high capacity Li Ion batteries for a long battery life The battery pack should last up to 3...

Страница 16: ...nue Anyway Mountz has tested the product with both the Home version and Professional version of XP You will not get this message if you are installing under Windows 2000 8 9 10 The initial screen will appear The Online LED should be green and Torque values are displayed on the right side of the screen For most users the only Tab that will be used is the Tool Tests The other Tabs are used for Trans...

Страница 17: ...yed on the right side of the screen For most users the only Tab that will be used is the Tool Tests The other Tabs are used for Transducer and Meter calibrations which should only be done by a certified calibration lab Image below is Torque Calibration screen 11 Image below is Meter Calibration screen ...

Страница 18: ...he Tool Tests tab Enter the relevant information on this screen such as the Test Name which is the name by which the test will be identified on the PTT Then enter all required information on the screen because once the tool test is activated on the PTT no changes can be made with the units tolerances or other information There is a field for comments to include user specific information The transd...

Страница 19: ...etrieves the results of a tool test after it has been run It will offer a change to add further notes to the test at this point The results will be saved in the PC in C PTT 2000 Tool Test Results Operator Folder where the Operator Folder will be the name of the Operator entered in the upper portion of the Tool Test Screen The results are stored in a csv file which can be opened in Excel or using a...

Страница 20: ... by a colon and then the number of data points collected Stop ends the test This is required as some users may want to collect 10 data or others may want to collect 25 points or more Clear Memory allows the test points to be cleared in the meter itself The operator will be asked for confirmation before the action takes place This function should only be used if the readings have been downloaded to...

Страница 21: ...lue the units of calibration and the transducer type Bridge or Brushless Once all required information is entered the Start button is clicked and its function changes to Continue as shown in figure below and procedural information will be given in the large text box A 2 point calibration should work well in all but exceptional circumstances Once the calibration is complete the calibration data wil...

Страница 22: ...Mountz for those that require it as an optional item P T T C a l i b r a t i o n D a t a L o g g i n g This program is used for graphing the data This feature can be utilized to evaluate and confirm torque specifications in both production or R D environments When the graph first appears you can enlarge it by clicking on the Zoom Box which is the square box made up of dashed lines just to the righ...

Страница 23: ...ta click the Graphing Icon that is on the spreadsheet the Torque and Angle will be plotted for you If you have your own Torque and Angle data you can just replace the example data with your own highlight it and click the Graphing Icon to plot it Angle Torque 6 3 64 21 3 62 36 3 63 137 4 34 153 3 99 167 4 05 182 4 23 196 4 30 211 4 13 225 3 74 240 3 99 254 4 32 269 4 52 283 4 31 298 4 01 313 4 18 3...

Страница 24: ...supplied CD to the computer installed the PTT 2000 Interface program and installed the USB drivers Locate the Bootloader folder on your computer as shown below Select the Bootloader folder Run the Setup exe file by right clicking on it and choosing Open After installation the program will be located in the Mountz folder when you select All Programs ...

Страница 25: ...ot load HEX file to device button Locate the HEX file update It will be in the Hex Files folder The file is named Mountz App hex This is the current application so there is no need to update but if the operator chooses he she can reload it to see how this feature works The process will take about 6 minutes as there is quite a lot of code in the product The status will be shown in the PTT Boot Load...

Отзывы: