Tolerance
Auto Clear
Angle Threshold
T+A Threshold
First Peak
Direction
The display holds the first detected torque peak applied. Before any torque is applied, the display show dashes in the
torque value area. Once peak is detected, the display will show the torque value. If a
second peak is detected then it will be displayed in the lower right (in small reverse video).
First Peak
First Peak Setup
Sensitivity:
Low
Min Peak: 5.0089 Lbf.in
Main
1. Press the
"Menu"
Button.
2. Use the
Scroll Down
key and select Setup by highlighting and pressing
Enter
.
3. Use the
Scroll Up or Down
key to toggle through the Setup options.
4. Press
Enter
key to select a setup setting for
First Peak
.
Filter
Units
Mode
Setup
Selecting First Peak
1. Use the
Scroll Up or Down
key to toggle through the sensitivity settings:
(Low, Medium & High)
2. Press the
Right
or
Left
key to move down to the Min Peak setting location
3. Use the
Scroll Up or Down
to set the Minimum Peak
4. Press
Enter
key once setting is complete.
5. Press
Main
button to return to main display screen.
Page 11
Note:
First Peak Sensitivity
This feature is provided for click wrenches. It is not relevant for Mountz Break-over
or Cam-over products such as MMTB, TB, TBIH, MTBN, TSN, TSP, and TSC.
High sensitivity will clearly display the "double" peak produced by a click wrench.
Low sensitivity will reduce "false peaks" caused by operator hesitation.
For most click wrenches try using Medium sensitivity this works for many types of
click wrenches such as the Mountz Titan and DM wrenches. As different types of
click wrenches differ in how quickly they move from a first peak (click) to a second
peak (over torque). Some experimentation, with sensitivity, may be required in
order to get repeatable results.