Output formats (see manual 6.6.1)
The output format determines the print output at measuring point scans and at
the memory output. Apart from the standard list format, with all measured
values given in a list, the column output format allows for a clear and
space-saving printout in
columns
. For this purpose, a printer will automatically
switch to the condensed character mode. Alarm lists during the measuring
cycle are not available for this format. The
spreadsheet format
is available to
further process measuring data by means of spreadsheet applications (see
manual 6.1).
Output channel and output format
are displayed in the dimension field. By
using the key
CH
▲
the following options can be successively selected. By
pressing
CH
▲
and holding (approx. 1s) a back switch is possible.
Ab. Designation
Output Format
U
Measured values interface
in a list
n U
Measured values interface
in columns
t U
Measured values interface
in spreadsheet format
a U
Alarms out memory to interface
in a list
S
Meas. values interface and memory
Meas. values one channel from memory
in a list
to analog output
n S
Meas. values interface and memory
in columns
t S
Meas. values interface and memory
in spreadsheet format
7.3.2 Measuring Cycle and Memory Activation
The measuring cycle is used for storing measured values, for cyclic averaging
(see manual 6.7.4) or for monitoring the measured values including alarm list
output in case of limiting values being exceeded. The display of the measuring
cycle has 6 digits (
hh:mm:ss
) in the function MEAS. CYCLE.
Function Meas. Cycle:
MEAS.
CYCLE
0 0:0 1:0 0
S
Example:
meas. cycle 1 min, memory activated ´
S
´
The
input
of the measuring cycle has 6 digits and format
hh:mm:ss
according
to 5.3. The
clearing
of the measuring cycle and, as a result, the switch-off of
the automatic scan can be achieved by using the keys
ENTER
,
CLEAR
.
The
memory activation
for all manual and cyclic measuring point scans (see
7.4.1) can be performed with the key
CH
▲
. An ´
S
´ is indicated on the display
following the measuring cycle.
Memory Activation:
MEAS. CYCLE
With
key:
CH
▲
ALMEMO
®
2290-8
35
Cyclic Measuring Point Scan
Содержание ALMEMO 2290-8
Страница 1: ...Operating Instructions Data Logger ALMEMO 2290 8 V2 1 04 12 2003 ALMEMO 2290 8 1...
Страница 50: ...50 ALMEMO 2290 8 Appendix...
Страница 51: ...ALMEMO 2290 8 51 Appendix...